  • Main application: Rapid Thermal Processing (RTP) system,
  • GasesAvailable gases: N2, O2 or forming gas (20% 5% of H2 + 80% 95% of N2). No vacuum.
  • 4" max wafer diameter, wafer carrier available.
  • Temperature /time limitations: max 60 min for dependent time limits
    • T < 400°C
    , max 30 min for
    • : 60 min
    • 400°C < T < 600°C
    , max 15 min for
    • : 30 min
    • 600°C < T < 800°C
    , max 5 min for
    • : 15 min
    • 800°C < T < 1000°C
    .
    • : 5 min
  • Max temperature ramp up rate is 150°C/s.
  • Up to 100 steps programmable processes.


Panel
borderColor#BFB8AF
bgColorwhite
titleColorBlack
borderWidth2
titleBGColor#DCE9EE
borderStylesolid
titleTool Overview

Responsible: Dmitry Suyatin Alexander Den Ouden

Location: Room Q158, EVA-Lab (Cleanroom Level 1)

License and Booking Required: Yes

Model and supplier: RTP-1200-100 from UniTemp GmbH.